Referencing Spectra in Topspin with BioTop (RECOMMENDED)
- If you don’t have the BioTop Topspin extension, install the spectrometer-related dependencies by running
expinstalland keeping the default values. - Execute
biotopto open the browser window. - Set the measurement conditions correctly: Labeling, Solvent, Tube, Temperature. You can also provide the FASTA sequence of the protein, and BioTop will automatically calculate the Molecular size. (However, the MW does not affect the axes shift during referencing.)
- Once ready, click Close and execute the command:
btproc biorefonly - You are all set. The axes should have shifted—check that they are correct.
Note
BioTop fails to reference 4D spectra and their 2D projections. In that case, you will use the 2D projections to align the 4D spectrum to the HSQC spectrum(s) during peak picking in POKY.
Calibrating (aka Referencing) Manually All Your 2D, 3D, 4D Spectra in Topspin
Calibration is the process of using a referenced spectrum (e.g.,
1H-15N HSQC) to reference another spectrum (e.g.,4D HCNH NOESY).
- Select the common axes between your 2D reference spectrum (for example, the
15N HSQC) and your 3D or 4D spectra.- Example: For a 15N-edited 3D NOESY →
F1: H, F2: N, F3: HN
and a 4D HCNH NOESY →F1: C, F2: HC, F3: N, F4: HN
you might chooseF2-F3(for the 3D) andF3-F4(for the 4D) to match the dimensions of the reference15N HSQC.
- Example: For a 15N-edited 3D NOESY →
- Create the N-HN positive projections to identify common peaks with strong signals (which will be used for calibration):
- For the 3D NOESY (15N-edited):
projplp 23 all all 23 - For the 4D HCNH NOESY:
projplp 34 all all 34This will create projections of the planes that correspond to the dimensions matching the
15N HSQC.
- For the 3D NOESY (15N-edited):
- Pick a reference peak in the
15N HSQC:- Go to the
15N HSQC, type.ppto manually pick a peak, select the chosen reference peak, typepeaks, and note down its coordinates in ppm.
- Go to the
- Go to the 3D spectrum and display its
F2-F3plane:- Click on the “23” icon to display the
F2-F3plane. - Type
.mdand select and drag the15N HSQCspectrum into the display. In the left lower panel, increase contour levels to see the strong signals clearly without showing too much noise. - Select the 3D spectrum and start navigating through the
F1axis by left-clicking and dragging on the “up-down arrow” icon. - While you navigate, watch for the reference peak from the
15N HSQCthat should start appearing as a strong signal in the overlaidF2-F3plane. - Once you see the strong signal of the selected reference peak, stop dragging, place your cursor at the peak center, and note its index(F1) coordinate (not the ppm coordinate).
- Click on the “23” icon to display the
- Return to the
F2-F3plane in the 3D NOESY:- Click the “return” icon to go back to the full
F2-F3plane view. - Click the “E” icon and select the
F2-F3visible plane. - Enter the index(F1) you noted down into the text box next to that plane.
- Click the “return” icon to go back to the full
- Zoom in on the reference peak:
- You should now see the reference peak with a strong signal.
- Zoom in and type
.cal, then left-click at the center of the reference peak. - In the pop-up window, enter the ppm coordinates you noted from the
15N HSQC. - Click OK. Both the
F2andF3axes of the 3D NOESY will be shifted accordingly.
- Repeat the procedure for the
34projection of the 4D NOESY to calibrate it.
Authors
- Thomas Evangelidis